1、Apply high accuracy MEMS sensor and digital technology, can measure positive, negative or differential pressure
2、It can measure system pressure of fan, blower, filter, furnace draft and orifice plate and can apply to various clean room, biological safety cabinet, clean bench, ducts collection, medical or pharmaceutical machine, etc.
3、High accuracy, excellent temperature compensation and electromagnetic interference (EMI) ability(industrial EMI level 3), good for applications in complex EMI environments of industrial systems or equipment
4、Multiple ranges and engineering units, optional LCD display with unit indication
5、Multiple outputs selection, over voltage and reverse polarity protection, high reliability
6、Very good performance/price ratio, with accuracy up to 0.5% and range as low as 25Pa
7、Function keys: zero reset, unit select, response time set, etc.
8、DP1C has relay for independent alarm or ON/OFF control
9、High protection rate up to IP65
Medium: non-combustible, non-corrosive air, insensitive to moisture, dust, condensation and oil
Working/Medium Temp.: -20~70°C
Temp. Compensation: 0~60°C
Range: 0~25/10000Pa, see ranges table
Working Pressure: overload 10xFS(<=1kPa)/8xFS(>1kPa)
burst 20xFS(<=1kPa)/10xFS(>1kPa)
Performance:
Accuracy %FS ±1.0% ±0.5%
Thermal effect %FS/°C(Zero/FS) ±0.05/0.08 ±0.03/0.04
Stability %FS/Year ±0.5 ±0.3
Response Time: 0.5/1/2/5 s, can be set by DIP switch
Transmitter output: 0~10V, 4~20mA(2 wires), 0~5V
Output Load: ≤500Ω(current), ≥2kΩ(voltage)
Communication: 1 RS485/Modbus, R/W enable, 9600 bps, terminal resistance settable
Display and Keys: Optional large LCD display with unit and backlight (N/A for 4~20mA)
Display resolution: 0.1 Pa (or equivalent), unit switchable
Display update time: <1s
Power: current 18.5~35VDC (RL=500Ω), 8.5~35VDC (RL=0Ω)
voltage 16~28VAC/16~35VDC;power consumption 1.5VA
Process Connection: 5mm ID tubing
Zero set: easy to reset by keys
Work Temp.: -20~70°C, 0~95%RH (Non cond.)
Storage Temperature: -30~85°C, 0~95%RH (Non cond.)
Medium Temperature: -40~100°C
Housing: fire retardant PC(UL94V-0), copper nozzle
Protection: IP65
Weight: 220g
Approval: CE, meet EN61326-1 for industrial equipment
Models
DP1C Differential Pressure Controller
Specifications
The controller is originated from the DP10x11, with no transmitter output but with 2 relay outputs, LCD display and 3 keys for displaying and setting control parameters. Therefore, most of its technical parameters are the same as those of DP10x11. The different parts are:
Relay: 2×SPST, 3A/30VDC, 3A/250VAC
Display and Keys: large LCD with unit , backlight and 3 keys
Models
Measuring Ranges
1. Set the 5 engineering units by DIP switch and the related LCD indicator will be on.
2. For zero center models, add “Z” at the end of the model. For example, DP1**Z, means the range
is -30-0-30Pa. Only ranges 1~6 have this selection.
3. If the measured value is -0.XXX, it will be displayed as -.XXX on the LCD.